Skip to product information
1 of 1

SKU:HC18316

In stock

CommaSil HPLC C18 Universal column, 100 x 2.1mm, 5um

CommaSil HPLC C18 Universal column, 100 x 2.1mm, 5um

Pack size: each, leadtime 5-7 days

Regular price £321.54 GBP
Regular price £430.00 GBP Sale price £321.54 GBP
Sale Sold out

CommaSil® C18 Universal HPLC Column

Click here for further information on the Commasil range

CommaSil® C18 HPLC Column is crafted from high-purity silica microspheres, ensuring a high level of homogeneity in particle size and pore structure. The activity of the silica microspheres is effectively mitigated through an advanced surface bonding and comprehensive capping process, rendering the columns suitable for separating analytes with varying hydrophobicity. This versatility positions CommaSil® C18 HPLC Columns as the prime choice among general-purpose columns. Furthermore, its exceptional chromatographic performance, superb selectivity, and reliable reproducibility make it the preferred option for method development under non-extreme chromatographic conditions.

Features

● General purpose columns boasting excellent column-to-column and lot-to-lot reproducibility.
● Exceptional selectivity and versatility across a wide range of applications.
● Universally applicable to pharmaceutical, environmental, and food safety testing fields.
● Equivalent to Waters Symmetry C18 series.

L-Tryptophan
Chromatographic column: CommaSil® C18 (4.6×150 mm, 3. 5μm)
Cat. #: HC18435
Detector: UV 280 nm
Flow rate: 0.8 ml/min
Column temperature: 35 ℃
Mobile phase: 10 mM ammonium acetate solution (pH 4.0):
methanol = 90:10
Chromatogram of L-Tryptophan

Olaquinol Metabolites
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: methanol, B: 1% aqueous formic acid solution
Elution mode: isocratic elution (A:B = 40:60)
Flow rate: 1.0 mL/min
Column temperature: 30 ℃
Injection volume: 40 μL
Detector: UV detector
Detection wavelength: 320 nm


Chromatogram of Olaquinol Metabolites

11 Synthetic Colorants
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile Phase: A: 20 mmoL/L ammonium acetate, B: methanol
Flow rate: 1.0 mL/min
Column temperature: 30 ℃
Injection volume: 10 μL
Detector: UV detector
Detection wavelength: 400-800 nm
Table 1. Gradient Elution Program
 
Time (min) A (%) B (%)
0.01 90 10
12.0 65 35
19.0 55 45
22.5 50 50
23.0 45 55
24.0 35 65
34.0 35 65
35.0 90 10
42.0 90 10



Chromatograms of 11 Synthetic Colorants

415 nm:                                      520 nm:                                        610 nm:
1. Tartrazine;                               1. New Red;                                  1. Indigotine;
2. Quinoline Yellow 1;                 2. Amaranthus Red;                     2. Brilliant Blue 1;
3. Quinoline Yellow 2;                 3. Carmine Cochineal;                  3. Brilliant Blue 2.
4. Quinoline Yellow 3;                 4. Sunset Yellow;
5. Quinoline Yellow 4.                 5. Allura Red;
                                                    6. Carmoisine;
                                                    7. Erythrosine.

Cefazolin Sodium
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: phosphate buffer, B: acetonitrile
Flow rate: 1.2 mL/min
Column temperature: 45 ℃
Injection volume: 10 μL
Detector: UV-Vis detector
Detection wavelength: 254 nm
Table 1. Gradient Elution Program
 
Time (min) A (%) B (%)
0 98 2
2.0 98 2
4.0 85 15
10 60 40
11.5 35 65
12 35 65
15 98 2


 
Chromatogram of Cefazolin Sodium

Abamectin
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile Phase: A: Methanol & Acetonitrile (50+50), B: Ultra-pure water
Elution mode: Gradient elution (A:B=90:10)
Flow rate: 1.0 mL/min
Column temperature: 40 ℃
Injection volume: 20 μL
Detector: UV detector


Chromatogram of Abamectin

6-Benzyladenine
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: methanol, B: 20 mmol/L ammonium acetate solution
Elution method: isocratic elution (A:B=50:50)
Flow rate: 1.0 mL/min
Column temperature: 30 ℃
Injection volume: 10 μL
Detector: UV detector
Detection wavelength: 267 nm


Chromatogram of 6-Benzyladenine

Benzo(a)pyrene
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: acetonitrile, B: water
Elution method: isocratic elution (A:B=88:12)
Flow rate: 1.0 mL/min
Column temperature: 25 ℃
Injection volume: 10 μL
Detector: FLD
Emission wavelength: 406 nm
Excitation wavelength: 384 nm


Chromatogram of Benzo(a)pyrene

Melamine
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: acetonitrile B: ion-pairing buffer reagent
Elution mode: isocratic elution (A:B=10:90)
Flow rate: 1.0 mL/min
Column temperature: 40 ℃
Injection volume: 20 μL
Detector: UV detector
Detection wavelength: 240 nm


  Chromatogram of Melamine

Ochratoxin A
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: acetonitrile - water - glacial acetic acid (96+102+2)
Flow rate: 1.0 mL/min
Column temperature: 35 ℃
Injection volume: 50 μL
Detection wavelength: 460 nm
Emission wavelength: 460 nm
Excitation wavelength: 333nm

Chromatogram of Ochratoxin A
View full details