SKU:HC18236
CommaSil HPLC C18 Universal column, 50 x 4.6mm, 5um
CommaSil HPLC C18 Universal column, 50 x 4.6mm, 5um
Pack size: each, leadtime 5-7 days
CommaSil® C18 Universal HPLC Column
Click here for further information on the Commasil range
CommaSil® C18 HPLC Column is crafted from high-purity silica microspheres, ensuring a high level of homogeneity in particle size and pore structure. The activity of the silica microspheres is effectively mitigated through an advanced surface bonding and comprehensive capping process, rendering the columns suitable for separating analytes with varying hydrophobicity. This versatility positions CommaSil® C18 HPLC Columns as the prime choice among general-purpose columns. Furthermore, its exceptional chromatographic performance, superb selectivity, and reliable reproducibility make it the preferred option for method development under non-extreme chromatographic conditions.
Features
● General purpose columns boasting excellent column-to-column and lot-to-lot reproducibility.● Exceptional selectivity and versatility across a wide range of applications.
● Universally applicable to pharmaceutical, environmental, and food safety testing fields.
● Equivelent to Waters Symmetry C18 series.
L-Tryptophan
Chromatographic column: CommaSil® C18 (4.6×150 mm, 3. 5μm)
Cat. #: HC18435
Detector: UV 280 nm
Flow rate: 0.8 ml/min
Column temperature: 35 ℃
Mobile phase: 10 mM ammonium acetate solution (pH 4.0):
methanol = 90:10
Chromatogram of L-Tryptophan
Olaquinol Metabolites
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: methanol, B: 1% aqueous formic acid solution
Elution mode: isocratic elution (A:B = 40:60)
Flow rate: 1.0 mL/min
Column temperature: 30 ℃
Injection volume: 40 μL
Detector: UV detector
Detection wavelength: 320 nm
Chromatogram of Olaquinol Metabolites
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile Phase: A: 20 mmoL/L ammonium acetate, B: methanol
Flow rate: 1.0 mL/min
Column temperature: 30 ℃
Injection volume: 10 μL
Detector: UV detector
Detection wavelength: 400-800 nm
Table 1. Gradient Elution Program
Time (min) | A (%) | B (%) |
0.01 | 90 | 10 |
12.0 | 65 | 35 |
19.0 | 55 | 45 |
22.5 | 50 | 50 |
23.0 | 45 | 55 |
24.0 | 35 | 65 |
34.0 | 35 | 65 |
35.0 | 90 | 10 |
42.0 | 90 | 10 |
Chromatograms of 11 Synthetic Colorants
415 nm: 520 nm: 610 nm:
1. Tartrazine; 1. New Red; 1. Indigotine;
2. Quinoline Yellow 1; 2. Amaranthus Red; 2. Brilliant Blue 1;
3. Quinoline Yellow 2; 3. Carmine Cochineal; 3. Brilliant Blue 2.
4. Quinoline Yellow 3; 4. Sunset Yellow;
5. Quinoline Yellow 4. 5. Allura Red;
6. Carmoisine;
7. Erythrosine.
Cefazolin Sodium
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: phosphate buffer, B: acetonitrile
Flow rate: 1.2 mL/min
Column temperature: 45 ℃
Injection volume: 10 μL
Detector: UV-Vis detector
Detection wavelength: 254 nm
Table 1. Gradient Elution Program
Time (min) | A (%) | B (%) |
0 | 98 | 2 |
2.0 | 98 | 2 |
4.0 | 85 | 15 |
10 | 60 | 40 |
11.5 | 35 | 65 |
12 | 35 | 65 |
15 | 98 | 2 |
Chromatogram of Cefazolin Sodium
Abamectin
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile Phase: A: Methanol & Acetonitrile (50+50), B: Ultra-pure water
Elution mode: Gradient elution (A:B=90:10)
Flow rate: 1.0 mL/min
Column temperature: 40 ℃
Injection volume: 20 μL
Detector: UV detector
Chromatogram of Abamectin
6-Benzyladenine
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: methanol, B: 20 mmol/L ammonium acetate solution
Elution method: isocratic elution (A:B=50:50)
Flow rate: 1.0 mL/min
Column temperature: 30 ℃
Injection volume: 10 μL
Detector: UV detector
Detection wavelength: 267 nm
Chromatogram of 6-Benzyladenine
Benzo(a)pyrene
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: acetonitrile, B: water
Elution method: isocratic elution (A:B=88:12)
Flow rate: 1.0 mL/min
Column temperature: 25 ℃
Injection volume: 10 μL
Detector: FLD
Emission wavelength: 406 nm
Excitation wavelength: 384 nm
Chromatogram of Benzo(a)pyrene
Melamine
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: A: acetonitrile B: ion-pairing buffer reagent
Elution mode: isocratic elution (A:B=10:90)
Flow rate: 1.0 mL/min
Column temperature: 40 ℃
Injection volume: 20 μL
Detector: UV detector
Detection wavelength: 240 nm
Chromatogram of Melamine
Ochratoxin A
Chromatographic column: CommaSil® C18 (4.6 mm×250 mm, 5 μm)
Cat#: HC18536
Mobile phase: acetonitrile - water - glacial acetic acid (96+102+2)
Flow rate: 1.0 mL/min
Column temperature: 35 ℃
Injection volume: 50 μL
Detection wavelength: 460 nm
Emission wavelength: 460 nm
Excitation wavelength: 333nm
Chromatogram of Ochratoxin A